Attenuation coefficient determination of printed ABS and PLA samples in diagnostic radiology standard beams

Texto completo
Autor(es):
Veneziani, G. R. ; Correa, E. L. ; Potiens, M. P. A. ; Campos, L. L. ; CostaFelix, RPB
Número total de Autores: 5
Tipo de documento: Artigo Científico
Fonte: WAKE CONFERENCE 2021; v. 733, p. 6-pg., 2016-01-01.
Resumo

IAEA code of practice TRS-457 states that standard phantoms should offer the same primary attenuation and scatter production as relevant body section of a representative patient. Material cost, availability and dimensional stability must also be considered. The goal of this study is to determine the attenuation coefficient of printed ABS and PLA samples in standard X-ray beams, verifying if phantoms printed with these materials could be an easier-handle substitute for PMMA, enabling the creation of different designs in an easier and cheaper way. Results show that PMMA presents higher attenuation coefficient, followed by PLA and ABS, which means that thinner PMMA layer creates higher radiation attenuation. (AU)

Processo FAPESP: 10/16437-0 - Caracterização e aplicações de materiais dosimétricos utilizando a luminescência opticamente estimulada e técnicas complementares
Beneficiário:Linda Viola Ehlin Caldas