SOL-GEL α-Al2O3 detectors: TL and OSL response to beta radiation beams

Texto completo
Autor(es):
Polo, Ivon Oramas ; Caldas, Linda V. E.
Número total de Autores: 2
Tipo de documento: Artigo Científico
Fonte: Radiation Physics and Chemistry; v. 167, p. 7-pg., 2020-02-01.
Resumo

The dosimetric properties of the SOL-GEL alpha-Al2O3 detectors with several different concentrations of impurities were studied in beta radiation beams. For the characterization of the detectors, the TL and OSL techniques were used. A TL spectrum was acquired. The detectors showed adequate OSL and TL dosimetric characteristics, such as good reproducibility, a glow curve with the peak at similar to 230 degrees C at 5 degrees C/s of heating rate, adequate OSL decay curve, good linearity of response and high phosphor sensitivity. A thermal treatment study was also performed. Moreover, the calibration factors and the lower detection limits were determined for beta radiation. These characteristics indicate the suitability of SOL-GEL alpha-Al2O3 detectors for the establishment of a transfer system or alternative/complementary method for beta radiation dosimetry. (AU)