Fully disposable microfluidic electrochemical device for detection of estrogen receptor alpha breast cancer biomarker

Texto completo
Autor(es):
Uliana, Carolina V. [1] ; Peverari, Camila R. [1] ; Afonso, Andre S. [2] ; Cominetti, Marcia R. [3] ; Faria, Ronaldo C. [1]
Número total de Autores: 5
Afiliação do(s) autor(es):
[1] Univ Fed Sao Carlos, Dept Chem, BR-13565905 Sao Carlos, SP - Brazil
[2] UFVJM, Inst Sci Engn & Technol, BR-39803371 Teofilo Otoni, MG - Brazil
[3] Univ Fed Sao Carlos, Dept Gerontol, BR-13565905 Sao Carlos, SP - Brazil
Número total de Afiliações: 3
Tipo de documento: Artigo Científico
Fonte: BIOSENSORS & BIOELECTRONICS; v. 99, p. 156-162, JAN 15 2018.
Citações Web of Science: 28
Resumo

A novel fully disposable microfluidic electrochemical array device (mu FED) was developed and successfully applied for detection of the biomarker estrogen receptor alpha (ERa). The FED was constructed using low-cost materials and an inexpensive home cutter printer enabled the manufacture of dozens of mu FEDs in less than 2 h, at a cost of less than US\$ 0.20 in material per device. The FED incorporates counter and reference electrodes and eight carbon-based working electrodes, which were modified with DNA sequences known as estrogen response elements (DNA-ERE), where ERa binds specifically. Paramagnetic particles heavily decorated with anti-ER alpha antibody and horseradish peroxidase (MP-Ab-HRP) were used to efficiently capture ERa from the sample solution. The ER alpha-MP-Ab-HRP bioconjugate formed was injected into the FED and incubated with the DNA-ERE-modified electrodes, followed by amperometric detection with application of -0.2 V vs. Ag vertical bar AgCl while a mixture of H2O2 and hydroquinone was injected into the microfluidic device. An ultralow limit of detection of 10.0 fg mL(-1) was obtained with the proposed method. The performance of the assay, in terms of sensitivity and reproducibility, was studied using undiluted calf serum, and excellent recoveries in the range of 94.7-108% were achieved for the detection of ERa in MCF-7 cell lysate. The FED system can be easily constructed and applied for multiplex biomarker detection, making the device an excellent cost-effective alternative for cancer diagnosis, especially in developing countries. (AU)

Processo FAPESP: 15/19890-1 - Desenvolvimento de métodos para diagnóstico de doenças utilizando partículas magnéticas e arranjo de imunossensores em sistema microfluídico descartável
Beneficiário:Ronaldo Censi Faria
Linha de fomento: Auxílio à Pesquisa - Regular
Processo FAPESP: 13/06750-1 - Desenvolvimento de imunossensores utilizando eletrodos descartáveis de carbono para aplicação na determinação de biomarcadores para câncer e bactérias patogênicas
Beneficiário:Ronaldo Censi Faria
Linha de fomento: Auxílio à Pesquisa - Regular
Processo FAPESP: 12/24079-2 - Utilização de eletrodos descartáveis no desenvolvimento de biossensores para detecção de biomarcadores tumorais
Beneficiário:Carolina Venturini Uliana
Linha de fomento: Bolsas no Brasil - Pós-Doutorado