InP Nanowire Biosensor with Tailored Biofunctionalization: Ultrasensitive and Highly Selective Disease Biomarker Detection

Texto completo
Autor(es):
Mostrar menos -
Janissen, Richard [1, 2] ; Sahoo, Prasana K. [1, 3] ; Santos, Clelton A. [4] ; da Silva, Aldeliane M. [1] ; von Zuben, Antonio A. G. [1] ; Souto, Denio E. P. [5] ; Costa, Alexandre D. T. [6] ; Celedon, Paola [7] ; Zanchin, Nilson I. T. [6] ; Almeida, Diogo B. [8, 1] ; Oliveira, Douglas S. [1] ; Kubota, Lauro T. [5] ; Cesar, Carlos L. [1] ; de Souza, Anete P. [4] ; Cotta, Monica A. [1]
Número total de Autores: 15
Afiliação do(s) autor(es):
[1] Univ Estadual Campinas, Gleb Wataghin Phys Inst, BR-13083859 Campinas, SP - Brazil
[2] Delft Univ Technol, Kavli Inst Nanosci, NL-2629 HZ Delft - Netherlands
[3] Univ S Florida, Dept Phys, Tampa, FL 33620 - USA
[4] Univ Estadual Campinas, Ctr Mol Biol & Genet Engn, Biol Inst, BR-13083875 Campinas, SP - Brazil
[5] Univ Estadual Campinas, Chem Inst, BR-13083970 Campinas, SP - Brazil
[6] Fundacao Oswaldo Cruz, Carlos Chagas Inst, BR-81310020 Curitiba, Parana - Brazil
[7] Mol Biol Inst Parana, BR-81310020 Curitiba, Parana - Brazil
[8] Univ Michigan, Dept Phys, Ann Arbor, MI 48109 - USA
Número total de Afiliações: 8
Tipo de documento: Artigo Científico
Fonte: Nano Letters; v. 17, n. 10, p. 5938-5949, OCT 2017.
Citações Web of Science: 16
Resumo

Electrically active field-effect transistors (FET) based biosensors are of paramount importance in life science applications, as they offer direct, fast, and highly sensitive label-free detection capabilities of several biomolecules of specific interest. In this work, we report a detailed investigation on surface functionalization and covalent immobilization of biomarkers using biocompatible ethanolamine and poly(ethylene glycol) derivate coatings, as compared to the conventional approaches using silica monoliths, in order to substantially increase both the sensitivity and molecular selectivity of nanowire-based FET biosensor platforms. Quantitative fluorescence, atomic and Kelvin probe force microscopy allowed detailed investigation of the homogeneity and density of immobilized biomarkers on different biofunctionalized surfaces. Significantly enhanced binding specificity, biomarker density, and target biomolecule capture efficiency were thus achieved for DNA as well as for proteins from pathogens. This optimized functionalization methodology was applied to InP nanowires that due to their low surface recombination rates were used as new active transducers for biosensors. The developed devices provide ultrahigh label-free detection sensitivities similar to 1 fM for specific DNA sequences, measured via the net change in device electrical resistance. Similar levels of ultrasensitive detection of similar to 6 fM were achieved for a Chagas Disease protein marker (IBMP8-1). The developed InP nanowire biosensor provides thus a qualified tool for detection of the chronic infection stage of this disease, leading to improved diagnosis and control of spread. These methodological developments are expected to substantially enhance the chemical robustness, diagnostic reliability, detection sensitivity, and biomarker selectivity for current and future biosensing devices. (AU)

Processo FAPESP: 13/02300-1 - Nanofios semicondutores: mecanismos de formação e aplicação em biossensores
Beneficiário:Mônica Alonso Cotta
Modalidade de apoio: Auxílio à Pesquisa - Regular
Processo FAPESP: 10/51748-7 - Análise estrutural e química de biofilmes de Xylella fastidiosa
Beneficiário:Mônica Alonso Cotta
Modalidade de apoio: Auxílio à Pesquisa - Regular
Processo FAPESP: 15/16611-4 - Nanofios semicondutores do tipo III-V: estudos de síntese voltados para aplicações em biologia
Beneficiário:Mônica Alonso Cotta
Modalidade de apoio: Auxílio à Pesquisa - Regular