Calibration and error investigation of large tipping bucket flow meters

Texto completo
Autor(es):
Schwamback, Dimaghi [1] ; Anache, Jamil A. A. [2] ; Wendland, Edson C. [1]
Número total de Autores: 3
Afiliação do(s) autor(es):
[1] Univ Sao Paulo, Sao Carlos Sch Engn EESC, Dept Hydraul Engn & Sanit, CxP 359, BR-13566590 Sao Carlos, SP - Brazil
[2] Univ Fed Mato Grosso do Sul, CxP 549, BR-79070900 Campo Grande, MS - Brazil
Número total de Afiliações: 2
Tipo de documento: Artigo Científico
Fonte: CATENA; v. 209, n. 2 FEB 2022.
Citações Web of Science: 0
Resumo

Inherent errors in tipping bucket flow meters may limit monitoring data reliability. In this work, we perform the static and dynamic calibration of four large tipping buckets, apply different regression curves and investigate the possible measurement error sources. The volumetric capacity (static calibration) of each piece of equipment was determined. They were tested (dynamic calibration) under ten flow intensities, ranging from low to high rainfall intensities (return period larger than 100 years). For each flow rate, the measurement was recorded during six time intervals (1, 2, 5, 10, 20 and 30 min) and four regression equations - linear, potential, T vs. 1/Q and quadratic - were tested. According to the static calibration, the equipment has a volumetric capacity of 11.63 mL (TB1), 64.16 mL (TB2), 139.86 mL (TB3) and 660.95 mL (TB4). When tested under different flow rates (dynamic calibration), underestimations were identified according to the size of the cavity: TB1 (3.31%), TB2 (5.75%), TB3 (9.33%) and TB4 (13.57%). Among the alternative curves, linear regression showed the best correlation (above 99%) with the monitored data. Using this method, the measurement errors were reduced to 1.35% (TB1), 0.04% (TB2), 3.18% (TB3) and 3.73% (TB4). We investigated how the different variables (tipping speed, cavity volumetric capacity and time interval of data collection) influenced the error. Errors follow a parabolic function of tipping velocity and a linear function of cavity volumetric capacity. The time interval of data collection interfered in the data sampled, however no statistical correlation was found. Among those variables, cavity size is the most important one. Given its low cost we aimed to minimize the inherent error in large tipping buckets flow meters and encourage its application, increasing in-situ collection of hydrological data. (AU)

Processo FAPESP: 19/24292-7 - Dinâmica da infiltração e escoamento superficial em vertentes com diferentes usos do solo e mudanças climáticas em área de Cerrado
Beneficiário:Dimaghi Schwamback
Modalidade de apoio: Bolsas no Brasil - Doutorado
Processo FAPESP: 15/03806-1 - Disponibilidade hídrica e riscos de contaminação em áreas de afloramento do Sistema Aquífero Guarani
Beneficiário:Edson Cezar Wendland
Modalidade de apoio: Auxílio à Pesquisa - Programa BIOEN - Temático