Relevant information of concomitants obtained from background signal using thermospray flame furnace atomic absorption spectrometry (TS-FF-AAS) and chemometric tools

Texto completo
Autor(es):
Miranda, Kelber [1] ; Maretti Silveira Bueno, Maria Izabel [2] ; Pereira-Filho, Edenir Rodrigues [1]
Número total de Autores: 3
Afiliação do(s) autor(es):
[1] Univ Fed Sao Carlos, Ctr Ciencias Exatas & Tecnol, Dept Quim, Grp Anal Instrumental Aplicada, BR-13565905 Sao Carlos, SP - Brazil
[2] Univ Estadual Campinas, Inst Quim, Dept Quim Analit, BR-13083970 Campinas, SP - Brazil
Número total de Afiliações: 2
Tipo de documento: Artigo Científico
Fonte: Journal of Analytical Atomic Spectrometry; v. 24, n. 3, p. 304-309, 2009.
Citações Web of Science: 11
Resumo

This study reports the use of background signals and multivariate calibration as a procedure to assess total concentration of concomitants in thermospray flame furnace atomic absorption spectrometry (TS-FF-AAS). TS-FF-AAS employs a metal tube atomizer above an air/acetylene flame and the atoms residence time is longer than that observed in FAAS (flame AAS). As a result, elevated background signals are obtained. The main purpose of this study is to make use of the background signals to predict the total concentration of concomitants. Background signals of 29 solutions containing Cd (25 mu g L(-1)), Pb (500 mu g L(-1)) and four concomitants (Ca, K, Mg and Na-total concentration values ranging from 0 to 5.52 g L(-1)) were recorded at two wavelengths: 228.8 nm for Cd and 283.3 nm for Pb. With the help of partial least squares, calibration models using either each wavelength alone or a combination of both were constructed to predict the total concentrations of concomitants. The best model showed root mean square error of validation (RMSEV) of 0.35 g L(-1) using background signals at 283.3 nm. A pharmaceutical sample containing 3 g L(-1) of the total concentration of Ca, K, Mg and Na was analyzed and 3% was the relative error. This sample and those used in the calibration/validation sets were also analyzed using X-ray fluorescence (XRF) and a good agreement was observed. This study extends the possibilities for the use of background signals in spectrometric techniques. (AU)

Processo FAPESP: 07/04515-4 - Avaliação da combinação da espectrometria de absorção atômica com determinação sequencial e a espectrometria de absorção atômica com tubo na chama e aerossol térmico
Beneficiário:Edenir Rodrigues Pereira Filho
Modalidade de apoio: Auxílio à Pesquisa - Regular
Processo FAPESP: 04/01970-4 - Melhoria do desempenho analítico em espectrometria de absorção atômica: estudo e aprimoramento de sistemas atomizadores
Beneficiário:Edenir Rodrigues Pereira Filho
Modalidade de apoio: Auxílio à Pesquisa - Jovens Pesquisadores