Busca avançada
Ano de início
Entree


Performance of DOFZ Diode as On-Line Gamma Dosimeter in Radiation Processing

Texto completo
Autor(es):
Camargo, F. ; Goncalves, J. A. C. ; Tuominen, E. ; Haerkoenen, J. ; Bueno, C. C. ; IEEE
Número total de Autores: 6
Tipo de documento: Artigo Científico
Fonte: 2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9; v. N/A, p. 2-pg., 2009-01-01.
Resumo

In this work, we report on results obtained with two rad-hard Diffusion Oxygenated Float Zone (DOFZ) silicon diodes as on-line gamma dosimeter in radiation processing. One device was not irradiated before using as a dosimeter, while the other received a gamma pre-dose of 700 kGy. The samples irradiation was performed using a 60Co source at a dose rate of 2.50 kGy/h from 5 kGy up 275 kGy. It was investigated the dosimetric response of these devices, operating in short-circuit current mode, with respect to the sensitivity dependence on dose and charge-dose linearity. Without any pre-dose, the diode exhibited a significant sensitivity decrease due to radiation induced point-defects in the crystal bulk. Conversely, the pre-irradiated device presented very stable current signals with a relative charge sensitivity of 0.9 mC/kGy. (AU)

Processo FAPESP: 03/12720-6 - Desenvolvimento de um espectrômetro de partículas e radiação eletromagnética baseada no uso de diodos especiais de Si
Beneficiário:Carmen Cecília Bueno
Modalidade de apoio: Auxílio à Pesquisa - Regular