Busca avançada
Ano de início
Entree
(Referência obtida automaticamente do Google Scholar, por meio da informação sobre o financiamento pela FAPESP e o número do processo correspondente, incluída na publicação pelos autores.)

An improved sampling approach in multi-pumping flow systems applied to the spectrophotometric determination of glucose and fructose in syrups

Texto completo
Autor(es):
Carneiro‚ J.M.T. ; Dias‚ A.C.B. ; Zagatto‚ E.A.G. ; Santos‚ J.L.M. ; Lima‚ J.L.F.C.
Número total de Autores: 5
Tipo de documento: Artigo Científico
Fonte: Analytica Chimica Acta; v. 531, n. 2, p. 279-284, 2005.
Resumo

A novel sampling approach in flow analysis exploiting the multi-pumping concept is proposed and applied to the spectrophotometric determination of glucose and fructose in invert sugar syrups. The samples are collected in gelatin capsules and brought to the laboratory. The capsules are placed directly into the dissolution chamber of a flow-batch system, therefore, the manual sample dissolution step is not required. All the involved analytical steps, including the capsule dissolution under acidic conditions, are efficiently controlled by using discretely actuated solenoid micro-pumps. Glucose and fructose determinations involve sugar oxidation by periodate and evaluation of the remaining periodate by monitoring the tri-iodide complex formed after addition of potassium iodide. The proposed system is very stable, handles about 50 samples per hour and requires 28.8 mug NaIO4 and 2.5 mg KI per determination. Precise results (R.S.D. < 2%) in agreement with HPLC are obtained. (C) 2004 Elsevier B.V. All rights reserved. (AU)

Processo FAPESP: 99/12124-7 - Emprego de novos materiais e diferentes ambientes reacionais para melhoria em sensibilidade e seletividade na análise de amostras biológicas e ambientais
Beneficiário:Elias Ayres Guidetti Zagatto
Modalidade de apoio: Auxílio à Pesquisa - Temático