Development of a multicommuted flow analysis procedure for simultaneous determination of sulfate and chloride in petroleum coke employing a homemade syringe pump and a LED-based photometer

Texto completo
Autor(es):
Santos, Felisberto G. [1] ; Pereira, Andreia C. [1] ; Cruz, Sandra M. [2] ; Bizzi, Cezar A. [2] ; Flores, Erico M. M. [2] ; Reis, Boaventura F. [1]
Número total de Autores: 6
Afiliação do(s) autor(es):
[1] Univ Sao Paulo, Ctr Energia Nucl Agr, BR-13400970 Piracicaba, SP - Brazil
[2] Univ Fed Santa Maria, Dept Quim, BR-97119900 Santa Maria, RS - Brazil
Número total de Afiliações: 2
Tipo de documento: Artigo Científico
Fonte: ANALYTICAL METHODS; v. 7, n. 11, p. 4769-4779, 2015.
Citações Web of Science: 9
Resumo

The current paper deals with procedures for simultaneous determination of sulfate and chloride in petroleum coke samples after microwave-induced combustion (MIC), which were implemented employing a multicommuted flow analysis process. The setup comprised a homemade syringe pump and two individual flow system manifolds, each coupled to a homemade LED-based photometer. Dedicated software enabled sampling and reading steps to be performed while displacing the syringe pistons forward and backward, allowing a sampling rate of 150 determinations per hour. The procedure for sulfate determination was based on the reaction of barium with sulfate and turbidimetric detection, while the procedure for chloride determination was based on the photometric mercury/thiocyanate method. In order to investigate the feasibility of both the equipment setup and the proposed procedures, samples of petroleum coke previously digested by microwave induced combustion (MIC) were analyzed. For accuracy assessment, samples were also analyzed employing ion chromatography and inductively coupled plasma optical emission spectrometry to determine chloride and sulfate, respectively. The paired t-test indicated no significant difference between the results for our method and those of the established methods, at a 95% confidence level. Linear responses (r = 0.999) were obtained with concentrations ranging from 10 to 700 mg L-1 for sulfate, and from 0.25 to 10 mg L-1 for chloride. (AU)

Processo FAPESP: 13/05728-2 - Desenvolvimento de dispositivo para propulsão de fluído e módulo de análise com habilidade para implementar os processos de análise por injeção sequencial e multicomutação em análise em fluxo
Beneficiário:Felisberto Gonçalves Santos Junior
Modalidade de apoio: Bolsas no Brasil - Doutorado
Processo FAPESP: 11/23498-9 - Sistemas avançados de análises químicas em fluxo: aplicações nas áreas tecnológica, agronômica e ambiental
Beneficiário:Elias Ayres Guidetti Zagatto
Modalidade de apoio: Auxílio à Pesquisa - Temático