Busca avançada
Ano de início
Entree
(Referência obtida automaticamente do Web of Science, por meio da informação sobre o financiamento pela FAPESP e o número do processo correspondente, incluída na publicação pelos autores.)

Characterization of Synchronous Vertical Dual View Inductively Coupled Plasma Optical Emission Spectrometer with Application for Water Analysis

Texto completo
Autor(es):
Amaral, Clarice D. B. ; Schiavo, Daniela ; Barros, Juan A. V. A. ; Machado, Raquel C. ; Nogueira, Ana Rita A. ; Nobrega, Joaquim A.
Número total de Autores: 6
Tipo de documento: Artigo Científico
Fonte: Analytical Letters; v. 49, n. 13, p. 2092-2098, 2016.
Citações Web of Science: 2
Resumo

The performance of an inductively coupled plasma optical emission spectrometer was evaluated using Ar, Ba, Mg, and Ni as test elements. Short-and long-term stabilities were excellent with relative standard deviations lower than 2%. The warm-up times in axial, radial, and synchronous vertical dual view (SVDV) modes were 2 min. The limits of detection and quantification for As, Ba, Cd, Cr, Cu, Hg, Ni, Pb, Sb, Se, and U were evaluated and the values were compatible with Brazilian and the US Environmental Protection Agency legislation for potable water. A standard reference material of trace elements in water (SRM 1643e) was used to verify the accuracy for As, Ba, Cd, Cr, Cu, Ni, Pb, Sb, and Se in axial and SVDV. Both configurations provided suitable accuracy and sensitivity. The elements were successfully determined in three bottled commercial drinking water and in two fortified samples using SVDV. (AU)

Processo FAPESP: 13/26672-5 - Desenvolvimento de procedimentos para a determinação de As, Ba, Cd, Cr, Pb e Si em amostras de fertilizantes minerais por espectrometria de emissão óptica com plasma induzido por micro-ondas (MIP OES)
Beneficiário:Raquel Cardoso Machado
Modalidade de apoio: Bolsas no Brasil - Doutorado