Gel and thermoluminescence dosimetry for dose verifications of a real anatomy simulated prostate conformal radiation treatment in the presence of metallic femoral prosthesis

Texto completo
Autor(es):
Rojas, Diana M. C. [1] ; Pavoni, Juliana F. [1] ; Arruda, Gustavo V. [2] ; Baffa, Oswaldo [1]
Número total de Autores: 4
Afiliação do(s) autor(es):
[1] Univ Sao Paulo, Fac Philosophy Sci & Letters, Dept Phys, Ribeirao Preto, SP - Brazil
[2] Univ Sao Paulo, Ribeirao Preto Med Sch, Ribeirao Preto, SP - Brazil
Número total de Afiliações: 2
Tipo de documento: Artigo Científico
Fonte: JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS; AUG 2021.
Citações Web of Science: 0
Resumo

This study aims to verify the dose delivery of prostate radiotherapy treatments in an adult pelvic phantom with two metallic hip and femur prosthesis using a four-field box technique. The prostate planned target volume (PTV) tridimensional (3D) dose distribution was evaluated using gel dosimetry, and thermoluminescent dosimeters (TLD) were used for point-dose measurements outside it. Both results were compared to the treatment planning system (TPS) dose calculation without using heterogeneity corrections to evaluate the influence of the metal in the dose distribution. MAGIC-f gel dosimeter (Methacrylic and Ascorbic acid in Gelatin Initiated by Copper with Formaldehyde) associated with magnetic resonance imaging was used. TLD were positioned at several points at the bone metal interface and the sacrum region. The comparison of the gel measured and the TPS calculated dose distributions were done using gamma analysis (3%/3 mm), and a pass rate of 93% was achieved. The TLD dose values at the bone-metal interface showed variations from the planned dose. However, at the sacrum region, where the beams did not intercept the prosthesis, there was a good agreement between TPS planning and TLD measurements. Our results show how the combination of 3D dosimetry and measurements at specific points in the phantom allowed a comprehensive view of the dose distribution and identified that care must also be paid to regions outside the PTV. (AU)

Processo FAPESP: 21/02254-6 - Aplicações de ferramentas de aprendizado de máquina em radioterapia
Beneficiário:Juliana Fernandes Pavoni
Modalidade de apoio: Auxílio à Pesquisa - Regular
Processo FAPESP: 13/07699-0 - Centro de Pesquisa, Inovação e Difusão em Neuromatemática - NeuroMat
Beneficiário:Oswaldo Baffa Filho
Modalidade de apoio: Auxílio à Pesquisa - Centros de Pesquisa, Inovação e Difusão - CEPIDs