Coronary 3D reconstruction using IVUS images only: a numeric phantom investigation

Texto completo
Autor(es):
Soares Matsumoto, Monica Mitiko ; Cardoso, Fernando Mitsuyama ; Lemos, Pedro Alves ; Furuie, Sergio Shiguemi ; D'hooge, J ; McAleavey, SA
Número total de Autores: 6
Tipo de documento: Artigo Científico
Fonte: MEDICAL IMAGING 2010: ULTRASONIC IMAGING, TOMOGRAPHY, AND THERAPY; v. 7629, p. 9-pg., 2010-01-01.
Resumo

Intravascular ultrasound (IVUS) examination offers a tomographic view of the vessel, having the catheter tip as reference. During examination, the catheter is pulled back with a constant speed (0.5 or 1.0 mm/s) and the ultrasound transducer captures cross-sectional slices of the coronary. Currently, 3D IVUS reconstruction is based on single-plane or biplane angiography together with IVUS images. In this work, we present a preliminary approach to reconstruct tridimensionally the catheter path and coronary, based only on IVUS sequence. We have proposed a numeric phantom framework: coronary simulation, catheter dynamic path simulation, IVUS acquisition, reconstruction and validation. Our method infers the catheter path inside the coronary, based on shortest path graph algorithm. To reconstruct morphology, we have associated the catheter path and the position of the frame with smoothness costs, and solved it as a minimization problem. In this experiment we have used three different morphologies (straight, one curve and two curves) and 60 random initializations each for the initial point and angle of catheter insertion. The results for the plane containing the centerline of the catheter were 95.8% true positive and 8.5% false positive rates. (AU)

Processo FAPESP: 07/53985-3 - Análise morfológica e dinâmica da coronária baseadas no processamento tridimensional de exames de ultrassom intravascular
Beneficiário:Mônica Mitiko Soares Matsumoto
Modalidade de apoio: Bolsas no Brasil - Doutorado