Busca avançada
Ano de início
Entree
(Referência obtida automaticamente do Web of Science, por meio da informação sobre o financiamento pela FAPESP e o número do processo correspondente, incluída na publicação pelos autores.)

Tuning the Selectivity of Ionic Liquid Stationary Phases for Enhanced Separation of Nonpolar Analytes in Kerosene Using Multidimensional Gas Chromatography

Texto completo
Autor(es):
Hantao, Leandro W. [1, 2] ; Najafi, Ali [1] ; Zhang, Cheng [1] ; Augusto, Pabio [2] ; Anderson, Jared L. [1, 3]
Número total de Autores: 5
Afiliação do(s) autor(es):
[1] Univ Toledo, Dept Chem, Toledo, OH 43606 - USA
[2] Univ Estadual Campinas, Inst Chem, BR-13084970 Campinas, SP - Brazil
[3] Univ Toledo, Sch Green Chem & Engn, Toledo, OH 43606 - USA
Número total de Afiliações: 3
Tipo de documento: Artigo Científico
Fonte: Analytical Chemistry; v. 86, n. 8, p. 3717-3721, APR 15 2014.
Citações Web of Science: 29
Resumo

In this study, a series of ionic liquids (ILs) are evaluated as stationary phases in comprehensive two-dimensional gas chromatography (GC X GC) for the separation of aliphatic hydrocarbons from kerosene. IL-based stationary phases were carefully designed to evaluate the role of cavity formation/dispersive interaction on the chromatographic retention of nonpolar analytes by GC X GC. The maximum allowable operating temperature (MAOT) of the IL-based columns was compared to that of commercial IL-based columns. Evaluation of the solvation characteristics of GC columns guided the selection of the best performing IL-based stationary phases for the resolution of aliphatic hydrocarbons, namely, trihexyl(tetradecyl)phosphonium tetrachloroferrate ({[}P-66614]{[}FeCl4]) and trihexyl(tetradecyl)phosphonium tris(pentafluoroethyl)trifluorophosphate ({[}P-66614]{[}PAP]) ILs. The best performing {[}P-6614]{[}FeCl4] IL-based column exhibited a MAOT of 320 degrees C, higher than the commercial SUPELCOWAX 10 (MAOT of 280 degrees C) and commercial IL-based columns (MAOT up to 300 degrees C). The structurally tuned {[}P-66614]{[}FeCl4] IL stationary phase exhibited improved separation of aliphatic hydrocarbons by GC X GC compared to the commercial columns examined (e.g., OV-1701, SUPELCOWAX 10, SLB-IL60, SLB-IL100, and SLB-IL111). (AU)

Processo FAPESP: 12/22375-3 - Fases Sortivas de líquido iônico: síntese, caracterização e aplicações em GCxGC
Beneficiário:Leandro Wang Hantao
Modalidade de apoio: Bolsas no Exterior - Estágio de Pesquisa - Doutorado