Evaluation of characteristic-to-total spectrum ratio: Comparison between experimental and a semi-empirical model

Full text
Author(s):
Gonzales, A. H. Lopez [1] ; Tomal, A. [2] ; Costa, P. R. [1]
Total Authors: 3
Affiliation:
[1] Univ Sao Paulo, Inst Phys, Lab Radiat Dosimetry & Med Phys, BR-66318 Sao Paulo, SP - Brazil
[2] Univ Estadual Campinas, Phys Inst Gleb Wataghin, BR-13083859 Campinas, SP - Brazil
Total Affiliations: 2
Document type: Journal article
Source: Applied Radiation and Isotopes; v. 100, p. 27-31, JUN 2015.
Web of Science Citations: 3
Abstract

Primary X-ray spectra were measured in the range of 80-150 kV in order to validate a computer program based on a semiempirical model. The ratio between the characteristic and total air Kerma was considered to compare computed results and experimental data. Results show that the experimental spectra have higher first HVL and mean energy than the calculated ones. The ratios between the characteristic and total air Kerma for calculated spectra are in good agreement with experimental results for all filtrations used. (C) 2015 Elsevier Ltd. All rights reserved. (AU)

FAPESP's process: 10/12237-7 - Experimento evaluation of the relation between air-Kerma and ambient equivalent dose for shielding calculation in adult and pediatric X-ray rooms