Advanced search
Start date
Betweenand


Dosimetric parameters and radiation tolerance of epitaxial diodes for diagnostic radiology and computed tomography beams

Full text
Author(s):
Gonsalves, Josemary A. C. ; Mangiarotti, Alessio ; Antonio, Patricia L. ; Caldas, Linda V. E. ; Bueno, Carmen C.
Total Authors: 5
Document type: Journal article
Source: Radiation Physics and Chemistry; v. 223, p. 8-pg., 2024-07-14.
Abstract

A custom-made EPI diode-based dosimetry system is thoroughly characterized for diagnostic radiology and computed tomography beams. The diode has a thin n-type epitaxial layer (50 mu m) grown on a thick (300 mu m) Czochralski silicon substrate. It operates as an online radiation dosimeter in the short-circuit current mode. In this case, the key dosimetric quantity is the dose rate, correlated with the output current from the diode exposed to radiation. The corresponding collected charge (the integral of the current signal) is proportional to the dose. Irradiations are performed with the Pantak-Seifert 160HS Isovolt X-ray generator previously standardized by Radcal RC6-RD and RC3-CT ionization chambers. The data gathered with all radiation quality beams confirm the linearity with dose and dose rate despite a slight energy dependence. Independently from the beam energy, the dosimetric parameters of repeatability ( <0.3%), long-term stability (0.4%/year), angular response ( <3%, +/- 5 degrees ), dose rate dependence ( <3%), and signal-to-noise ratio (>= 4900) fully adhere to the IEC 61674 recommendations. Compliance with the accumulated dose stability requirement (1.0%/40 Gy) is almost achieved with the pristine diode and effectively accomplished through radiation conditioning with Co-60 gamma rays. Under the latter condition, the lifespan of the diode can easily reach 15 kGy, assuring the high reusability of this diode for diagnostic radiology and computed tomography dosimetry before requiring recalibrations. (AU)

FAPESP's process: 18/05982-0 - R&D&I in radiation metrology in the health area
Grantee:Linda Viola Ehlin Caldas
Support Opportunities: Research Projects - Thematic Grants
FAPESP's process: 22/13430-2 - Development and implementation of a procedure based on the phototransfer phenomenon for the characterization of dosimetric materials with application in radiation dosimetry
Grantee:Patrícia de Lara Antonio
Support Opportunities: Scholarships in Brazil - Technical Training Program - Technical Training