Torque loss under mechanical cycling of long-span zirconia and titanium-cemented and screw-retained implant-supported CAD/CAM frameworks

Full text
Author(s):
Gomes, Erica Alves [1] ; Tiossi, Rodrigo [1] ; Lapria Faria, Adriana Claudia [1] ; Rodrigues, Renata C. S. [1] ; Ribeiro, Ricardo Faria [1]
Total Authors: 5
Affiliation:
[1] Univ Sao Paulo, Dent Sch Ribeirao Preto, Dept Dent Mat & Prosthodont, BR-14049 Ribeirao Preto - Brazil
Total Affiliations: 1
Document type: Journal article
Source: Clinical Oral Implants Research; v. 25, n. 12, p. 1395-1402, DEC 2014.
Web of Science Citations: 3
Abstract

Objectives: This study evaluated the screw joint stability after cyclic loading of implant-supported titanium and zirconia CAD/CAM frameworks for fixed dental prostheses (FDPs) with different retention methods. Material and methods: Twenty four one-piece frameworks supported by six threaded implants placed in the maxilla were fabricated using a CAD/ CAM technique (NeoShape). Dry-pressed porcelain crowns were luted to the frameworks to standardize the specimens. The specimens were then divided into four groups (n = 6) according to framework material (titanium or zirconia) and retention method for the prosthesis (cement-or screw-retained): G1, Ti-cemented; G2, Ti-screw-retained; G3, Zr-cemented; and G4, Zr-screw-retained. A digital torque ratchet was used to assess the initial preload removal torque. Torque was then reapplied and the specimens were submitted to a 200 N cyclic load, at a frequency of 2 Hz, underwater in controlled temperature of 37 degrees, and for 1 x 10(6) cycles. An opposing lower dental arch was fabricated using bis-acrylic resin to simulate occlusal contacts in centric. After cyclic loading, postload removal torque was measured. Preload and postload torque loss was expressed as a percentage of the initial load. Data were submitted to a linear mixed-effects model for statistical significance (alpha = 0.05) to evaluate the effect of cyclic loading in the screw torque loss used with frameworks of different materials and retention methods. Results: Significant screw torque loss (%) was found for the tested groups (before/after cyclic loading, respectively): G1 (39.77/61.83), G2 (37.57/50.96), G3 (34.87/54.10), and G4 (47.56/73.50) (P < 0.05). Conclusions: The screw removal torque was significantly reduced for all groups in this study after cyclic loading the specimens. Screw-retained zirconia specimens presented the highest torque loss before and after the cyclic loadings compared with the other specimens that were tested. (AU)

FAPESP's process: 10/02218-5 - Influence of mechanical cycling on loosening of retention screw in cemented and screwed fixed partial dentures fabricated with titanium and zirconia cad/cam system. In vitro analysis and non-linear analysis by 3D-FEA
Grantee:Ricardo Faria Ribeiro
Support Opportunities: Regular Research Grants
FAPESP's process: 10/00124-3 - Influence of mechanical cycling on loosening of retention screw in cemented and screwed fixed partial dentures fabricated with titanium and zirconia cad/cam system. In vitro analysis and non-linear analysis by 3D-FEA.
Grantee:Erica Alves Gomes
Support Opportunities: Scholarships in Brazil - Post-Doctoral