Multi-energy calibration (MEC) applied to laser-induced breakdown spectroscopy (LIBS)

Full text
Author(s):
Babos, Diego Victor [1] ; Virgilio, Alex [1] ; Costa, Vinicius Camara [1] ; Donati, George Luis [2] ; Pereira-Filho, Edenir Rodrigues [1]
Total Authors: 5
Affiliation:
[1] Univ Fed Sao Carlos, Dept Chem, Grp Appl Instrumental Anal, POB 676, BR-13565905 Sao Carlos, SP - Brazil
[2] Wake Forest Univ, Dept Chem, Salem Hall, Box 7486, Winston Salem, NC 21709 - USA
Total Affiliations: 2
Document type: Journal article
Source: Journal of Analytical Atomic Spectrometry; v. 33, n. 10, p. 1753-1762, OCT 1 2018.
Web of Science Citations: 14
Abstract

In this study, the multi-energy calibration (MEC) method is applied to solid sample analysis by laser-induced breakdown spectroscopy (LIBS). The MEC method uses only two calibration standards and several atomic emission wavelengths with different sensitivities to determine the analyte concentration in the sample. Both calibration mixtures are prepared using the same amount of sample, which contributes to minimizing matrix effects. The MEC-LIBS method was used to determine Ca, Cu, Fe, Mn and Zn in complex-matrix samples of cattle mineral supplements with an adequate sample throughput (15 samples per h). The limits of detection (LODs) calculated for these elements in the solid sample were 11 g kg(-1), 45, 188, 53 and 44 mg kg(-1), respectively. For Cu, Mn and Zn, these LODs were approximately 10, 10, and 20 times lower than the minimum level required in a commercial product by the Brazilian legislation. The method's accuracy was evaluated by analyzing five reference materials (RM) of cattle mineral supplements, and by comparing values determined in six commercial samples with results from ICP-OES determinations. No statistically significant difference was observed between MEC-LIBS and RM or ICP-OES-determined values at the 95% confidence level (Student's t-test, n = 3). LIBS is an inherently atomic-emission-rich technique, which suffers from severe matrix effects. MEC is an efficient matrix-matching method, for which the larger the number of analytical signal sources for an analyte the more precise the results. Therefore, MEC is a highly compatible, complementary approach to LIBS calibration. (AU)

FAPESP's process: 14/18393-1 - Evaluation of gradient flow analysis as calibration strategy for spectroanalytical techniques with plasma source
Grantee:Alex Virgilio
Support Opportunities: Scholarships in Brazil - Post-Doctoral
FAPESP's process: 16/01513-0 - Laser-Induced Breakdown Spectroscopy (LIBS) application in the analyses of electronic waste (printed circuit boards and polymers), food and difficult preparation liquid samples
Grantee:Edenir Rodrigues Pereira Filho
Support Opportunities: Regular Research Grants