Advanced search
Start date
Betweenand
(Reference retrieved automatically from Web of Science through information on FAPESP grant and its corresponding number as mentioned in the publication by the authors.)

Determination of parabens in urine samples by microextraction using packed sorbent and ultra-performance liquid chromatography coupled to tandem mass spectrometry

Full text
Author(s):
Jardim, Valeria Cristina [1] ; Melo, Lidervan de Paula [1] ; Domingues, Diego Soares [1] ; Costa Queiroz, Maria Eugenia [1]
Total Authors: 4
Affiliation:
[1] Univ Sao Paulo, Fac Filosofia Ciencias & Letras Ribeirao Preto, Dept Quim, BR-14040901 Ribeirao Preto, SP - Brazil
Total Affiliations: 1
Document type: Journal article
Source: JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES; v. 974, p. 35-41, JAN 1 2015.
Web of Science Citations: 17
Abstract

A simple, sensitive, and selective method using ultra-performance liquid chromatography coupled to tandem mass spectrometry (UPLC-MS/MS) was developed and validated for simultaneous determination of parabens {[}methyl paraben (MeP), ethyl paraben (EtP), propyl paraben (PrP), butyl paraben (BuP), and benzyl paraben (BzP)] in human urine samples. After microextraction by packed sorbent (MEPS) using a C18 phase, the parabens were separated on a Kinetex C18 column (100 mm x 2.1 mm x 1.7 mu m) within 4.6 min using isocratic elution. These compounds were detected on a triple quadrupole tandem mass spectrometer using the multiple reactions monitoring (MRM) mode via an electrospray ionization source operating in the negative ionization mode. Important factors that influence MEPS performance were evaluated, such as the sample pH, draw-eject sample volume, clean-up step, and desorption conditions. The proposed MEPS/UPLC-MS/MS method presented a linear range from 0.5 ngmL(-1) (limit of quantification - LOQ) to 50 ngmL(-1), and interassay precision with coefficients of variation lower than 15%, and relative standard error values of the accuracy ranged from -8.8% to 15%. The MEPS/UPLC-MS/MS method was applied successfully to determine parabens in urine samples from 30 postpartum volunteers, enabling assessment of human exposure to these compounds. (C) 2014 Elsevier B.V. All rights reserved. (AU)

FAPESP's process: 12/10705-9 - Development of selective phases for in-tube SPME/LC-MS/MS analysis of parabens and drugs in biological samples
Grantee:Maria Eugênia Queiroz Nassur
Support Opportunities: Regular Research Grants