Advanced search
Start date
Betweenand


Synchrotron light source focused X-ray detection with LGADs, AC-LGADs and TI-LGADs

Full text
Author(s):
Show less -
Molnar, A. ; Davis, M. ; Oregan, G. ; Beringer, S. ; Zhao, Y. ; Mazza, S. M. ; Tiernan, A. ; Ott, J. ; Sadrozinski, H. F-W ; Seiden, A. ; Schumm, B. ; McKinney-Martinez, F. ; Bisht, A. ; Centis-Vignali, M. ; Paternoster, G. ; Boscardin, M.
Total Authors: 16
Document type: Journal article
Source: Journal of Instrumentation; v. 20, n. 7, p. 11-pg., 2025-07-01.
Abstract

The response of Low Gain Avalanche Diodes (LGADs), a type of thin silicon detector with internal gain, to X-rays of energies between 6-16 keV was characterized at the Stanford Synchrotron Radiation Lightsource (SSRL). The utilized beamline at SSRL was 7-2, with a nominal beam size of 30 mu m, repetition rate of 500 MHz, and with an energy dispersion Delta E/E of 10-4. Multi-channel LGADs, AC-LGADs, and TI-LGADs of different thicknesses and gain layer configurations from Hamamatsu Photonics (HPK) and Fondazione Bruno Kessler (FBK) were tested. The sensors were read out with a discrete component board and digitized with a fast oscilloscope or a CAEN fast digitizer. The devices' energy response, energy resolution, and time resolution were measured as a function of X-ray energy and position. The charge collection and multiplication mechanism were simulated using TCAD Sentaurus, and the results were compared with the collected data. (AU)

FAPESP's process: 20/04867-2 - High energy physics and instrumentation with the LHC-CERN
Grantee:Marcelo Gameiro Munhoz
Support Opportunities: Special Projects