Fatigue Failure of External Hexagon Connections on Cemented Implant-Supported Crowns

Full text
Author(s):
Barbosa, Joao Malta [1, 2] ; da Rocha, Daniel Navarro [3] ; Hirata, Ronaldo [4] ; Freitas, Gileade [5] ; Bonfante, Estevam A. [6] ; Coelho, Paulo G. [7]
Total Authors: 6
Affiliation:
[1] NYU, Dept Prosthodont, New York, NY - USA
[2] NYU, Coll Dent, Dept Biomat & Biomimet, 433 1st Ave, Room 844, New York, NY 10010 - USA
[3] Mil Inst Engn, Dept Mat Sci, Rio De Janeiro, RJ - Brazil
[4] NYU, Dept Biomat & Biomimet, New York, NY - USA
[5] Univ Sao Paulo, Ribeirao Preto Dent Sch, Dept Oral & Maxillofacial Surg, Ribeirao Preto, SP - Brazil
[6] Univ Sao Paulo, Bauru Sch Dent, Dept Prosthodont & Periodontol, Bauru, SP - Brazil
[7] NYU, Tandon Sch Engn, Hansjorg Wyss Dept Plast Surg, Mech & Aerosp Engn, Dept Biomat & Biomimet, Langone Med Ctr, New York, NY - USA
Total Affiliations: 7
Document type: Journal article
Source: IMPLANT DENTISTRY; v. 27, n. 2, p. 165-170, APR 2018.
Web of Science Citations: 1
Abstract

Purpose: To evaluate the probability of survival and failure modes of different external hexagon connection systems restored with anterior cement-retained single-unit crowns. The postulated null hypothesis was that there would be no differences under accelerated life testing. Materials and Methods: Fiftyfour external hexagon dental implants (similar to 4 mm diameter) were used for single cement-retained crown replacement and divided into 3 groups: (3i) Full OSSEOTITE, Biomet 3i (n = 18); (OL) OEX P4, Osseolife Implants (n = 18); and (IL) Unihex, Intra-Lock International (n = 18). Abutments were torqued to the implants, and maxillary central incisor crowns were cemented and subjected to step-stress-accelerated life testing in water. Use-level probability Weibull curves and probability of survival for a mission of 100,000 cycles at 200 N (95% 2-sided confidence intervals) were calculated. Stereo and scanning electron microscopes were used for failure inspection. Results: The beta values for 3i, OL, and IL (1.60, 1.69, and 1.23, respectively) indicated that fatigue accelerated the failure of the 3 groups. Reliability for the 3i and OL (41% and 68%, respectively) was not different between each other, but both were significantly lower than IL group (98%). Abutment screw fracture was the failure mode consistently observed in all groups. Conclusion: Because the reliability was significantly different between the 3 groups, our postulated null hypothesis was rejected. (AU)

FAPESP's process: 16/18818-8 - Multi-User Equipment approved in grant 2012/19078-7: ElectroPuls E3000 Linear-Torsion All-Electric dynamic test instrument package
Grantee:Estevam Augusto Bonfante
Support Opportunities: Multi-user Equipment Program
FAPESP's process: 12/19078-7 - Development and processing of Al2O3-ZrO2 translucent composite for CAD/CAM monolithic prostheses: the effect of aging on biaxial flexural strength
Grantee:Estevam Augusto Bonfante
Support Opportunities: Research Grants - Young Investigators Grants